7�4 System Workstation Power Loss; 7�5 System Workstation Becomes Non-Operational; Corrective Action - Boston Scientific LithoVue Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 46
7� 4 SyStEM WorKStAtIon PoWEr LoSS
cAUtIon: In the event of a loss of System Workstation power, do not AdVAncE tHE
UrEtEroScoPE or Any ProcEdUrAL dEVIcES In tHE WorKInG cHAnnEL WItHoUt
VIEWInG tHE LIVE IMAGE�
Follow these steps to recover from an electrical power loss:
1. Press the Power Button to restart the System Workstation.
2. If the System Workstation fails to power on after the Power Button is pressed, verify that the unit
is plugged into the hospital's electrical outlet, that the Power Cord is plugged into the System
Workstation, AND that the Power Transformer switch is turned ON (the green LED on the Power
Transformer will be lit).
3. Press the Power button to restart the System Workstation.
4. If the System Workstation fails to start, have biomedical personnel verify correct functioning of
the hospital's electrical outlet and restart the System Workstation.
5. If the System Workstation fails to start after the hospital's electrical outlet has been verified,
contact Boston Scientific using the information found in Section 8 "Service and Warranty" for
System Workstation service.
7� 5 SyStEM WorKStAtIon BEcoMES non-oPErAtIonAL
cAUtIon: In the event of a system lockup, do not AdVAncE tHE UrEtEroScoPE or Any
ProcEdUrAL dEVIcES In tHE WorKInG cHAnnEL WItHoUt VIEWInG tHE LIVE IMAGE�
Follow these steps to recover if the System Workstation becomes non-operational:
1. Press the Power button until the System Workstation shuts down.
2. Press the Power button to restart the System Workstation.
3. If the System Workstation fails to start, contact Boston Scientific using the information found in
Section 8 "Service and Warranty" for System Workstation service.
table 7�1 LithoVue™ System Workstation troubleshooting chart
User Message
LithoVue Flexscope
Disconnected icon is
displayed in video field.
User's Manual
Possible cause
Flexscope Cable is not properly
or fully inserted into the cable
receptacle on the System
Workstation.
LithoVue Flexscope is broken
or defective.
Connection of
unapproved device.

corrective Action

Remove the LithoVue Flexscope. Ensure
the connection terminals are dry and
clean. Ensure the plug on the Flexscope
Cable is properly aligned with the key on
the Workstation Receptacle. Ensure the
plug is fully inserted into the receptacle.
If User Message remains after a repeated
connection, replace the LithoVue Flexscope.
Remove the LithoVue Flexscope. Ensure
the connection terminals are dry and
clean. Ensure the plug on the Flexscope
Cable is properly aligned with the key on
the Workstation Receptacle. Ensure the
plug is fully inserted into the receptacle.
If User Message remains after a repeated
connection, replace the LithoVue Flexscope.
Replace unapproved device with a
LithoVue Flexscope.
39
Black (K) ∆E ≤5.0

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

M0067911000

Tabla de contenido