Warnings/Error Signals - Ottobock Meridium 1B1 Instrucciones De Uso

Ocultar thumbs Ver también para Meridium 1B1:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 145
Beep signal
Vibration
signal
2x short
2x short
3x short
3x short

13.2.2 Warnings/error signals

Error during use
Beep signal
10x long
Beep signal
10 x short
5 long beep signals
when malfunction
occurs
30 x long
Additional action per­
formed
Weight placed on prosthetic
leg and leg kept still for 1
second
Weight placed on prosthetic
leg and leg kept still for 1
second
Vibration signal
3x long
5x long
10x long
Vibration signal
1 x short at interval of
approx. 5 seconds
10 x long
5 x short at interval of
5 minutes
Intermittent for
5 minutes
Event
Switching to MyMode 1 (mode 2)
carried out.
Switching to MyMode 2 (mode 3)
carried out.
Event
Battery charge level under 25%
Battery charge level under 10%
Battery charge level under 5%
Event
Overheated hydraulic
unit
Battery charge level
under 5 %
Error of moderate
severity:
e.g. a sensor is not
operational
Switch to safety
mode.
Active safety mode
Severe error
e.g. failure of valve
drives
Possibly no switching
into safety mode.
Required action
Reduce activity
Charge battery
Restricted walking pos­
sible.
The product must be
inspected immediately by
a prosthetist.
Restricted walking pos­
sible.
The product must be
inspected immediately by
a prosthetist.
Restricted walking pos­
sible.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger.
If the error persists, it is
no longer permissible to
use the product. The
product must be inspec­
ted immediately by a
prosthetist.
Ottobock | 67

Publicidad

Tabla de contenido
loading

Tabla de contenido