16.2.2 Warnings/error signals
Error during use
Beep signal
–
–
–
5 x long
10 x long
30 x long
–
100
Vibration signal
1 x long at intervals of
approx. 5 seconds (if
mute mode (silent
mode) is activated,
this signal is not gen
erated)
3 x long
5 x long
5 x long repeated
every 60 seconds
10 x long
1 x long, 1 x short
repeated every
3 seconds
Continuous
Event
Overheated hydraulic
unit
Charge level under
25%
Charge level under
10%
Medium error (see
Page 92)
e.g. a sensor is not
operational
Battery charge level
5%
After the beep and
vibration signals, the
product switches to
empty battery mode
and then switches
off.
Severe error/indic
ation of safety
mode activation
(see Page 92)
e.g. one or more
sensors are not oper
ational
Total failure
Electronic control no
longer possible.
Safety mode active or
undetermined valve
state. Undetermined
product behaviour.
Required action
Reduce activity.
Charge battery soon.
Remaining operating time
approx. 24 hours
Charge battery soon
Remaining operating time
approx. 6 hours
Walking possible with
restrictions. Please note
the change in flexion res
istance.
The product must be
inspected immediately by
an O&P professional.
Charge the battery.
Attempt to reset this error
by connecting/discon
necting the battery char
ger.
If the error persists, use
of the product is prohib
ited. The product must be
inspected immediately by
an O&P professional.
Attempt to reset this error
by connecting/discon
necting the battery char
ger.
If the error persists, use
of the product is prohib
ited. The product must be
inspected immediately by
an O&P professional.