6. 3D TOMOGRAPHY (CBCT)

Only for 3D machines
The CBCT exam consists in the three-dimensional reconstruction of an X-rayed anatomical region, which produces
both two-dimensional views and three-dimensional representations that are generated by a software on a workstation
(PC).
Instructions about image processing can be found in the iRYS software user manual.
Remember to change the disposable protective covers before positioning a new patient.
Before positioning the patient, make sure all metal items worn have been taken off such as eyeglasses,
movable prosthesis, earrings and any other removable metal items present at head level. If an apron is used
for protection against radiation, make sure the patient's neck is not covered. If it is, an area will not be X-
rayed.

6.1. SELECTING THE EXAMINATION FROM THE CONTROL CONSOLE

The 3D exam can only be performed if the X-ray device is connected to a PC where the iRYS software is
installed.
1) Turn the PC on and run the iRYS program
2) Upon the first start of each day, the daily control procedure (Daily Check) has to be completed before any 3D exam
can be started. This daily check is a service procedure with X-ray emission that has to be performed WITHOUT
THE PATIENT.
3) For instructions, please consult the "Acquisition Operations" section of the iRYS manual.
The patient must not access the X-ray device during the exam selection and the device reset operations.
Position the patient only after the device reset has been carried out according to the procedure described in
this paragraph.