Tabelle 2-10: Entity-Werte, aktiver Eingang, digitaler I/O
Belegung (Ort)
U
o
A_IN1+ (J2-8)
5.88 V
A_IN1- (J2-7)
A_IN2+ (J2-6)
5.88 V
A_IN2- (J2-5)
A_IN3+ (J2-4)
5.88 V
A_IN3- (J2-3)
A_IN4+ (J2-2)
5.88 V
A_IN4- (J2-1)
Tabelle 2-11: Entity-Werte, passiver Eingang, digitaler I/O
Belegung (Ort)
U
/ V
i
P_IN1+ (J2-8)
30 V
P_IN1- (J2-7)
P_IN2+ (J2-6)
30 V
P_IN2- (J2-5)
P_IN3+ (J2-4)
30 V
P_IN3- (J2-3)
P_IN4+ (J2-2)
30 V
P_IN4- (J2-1)
Tabelle 2-12: Entity-Werte, aktiver Ausgang, digitaler I/O
Belegung (Ort)
U
/
o
A_OUT1+ (J3-12)
12.6 V
A_OUT1- (J3-11)
A_OUT2+ (J3-10)
12.6 V
A_OUT2- (J3-9)
A_OUT3+ (J3-8)
12.6 V
A_OUT3- (J3-7)
A_OUT4+ (J3-6)
12.6 V
A_OUT4- (J3-5)
A_OUT5+ (J3-4)
12.6 V
A_OUT5- (J3-3)
A_OUT6+ (J3-2)
12.6 V
/
I
/ I
P
V
o
sc
oc
2 mA
2.94 mW
Erde
2 mA
2.94 mW
Erde
2 mA
2.94 mW
Erde
2 mA
2.94 mW
Erde
I
/ I
P
max
i
max
50 mA
375 mW
Erde
50 mA
375 mW
Erde
50 mA
375 mW
Erde
50 mA
375 mW
Erde
Io / Isc
P
Voc
o / Pt
92 mA
627 mW
Erde
92 mA
627 mW
Erde
92 mA
627 mW
Erde
92 mA
627 mW
Erde
92 mA
627 mW
Erde
92 mA
627 mW
METTLER TOLEDO IND560x Terminal Installation Manual
C
/ C
L
/ L
a
o
o
o
100 nF
100 H
100 nF
100 H
100 nF
100 H
100 nF
100 H
C
L
i
i
i
10 nF
10 µH
10 nF
10 µH
10 nF
10 µH
10 nF
10 µH
C
/ Ca
L
/ La
o
o
100 nF
400 µH
100 nF
400 µH
100 nF
400 µH
100 nF
400 µH
100 nF
400 µH
100 nF
400 µH
a
2-7