Capacity Label Instructions; Sealing The Enclosure; Sealing The Panel-Mount Enclosure - Mettler Toledo IND560x Manual De Instalación

Ocultar thumbs Ver también para IND560x:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 93
3.11.

Capacity Label Instructions

The regulations in some locations require that the scale capacity and increment be shown on the
front of the terminal, near the display. To satisfy this requirement, a blue capacity label is included
with the terminal that must be completed and adhered to the front overlay.
The capacity label (shown in Figure 3-50) provides space for the Max, min, and e information for
each range or interval for which the scale is programmed. If only one or two ranges are used, the
unused portion of the label may be cut off with scissors. Written information must be legible and a
minimum of 2mm or 0.08 in. in height. A permanent marker should be used for this information.
Clean any oil or other contaminants from the area of the overlay shown in Figure 3-51 where the
capacity label will be added. Peel the backing from the label and adhere it to the overlay in the
location shown in Figure 3-51, or another location acceptable to the local regulations.
3.12.

Sealing the Enclosure

When the IND560x terminal is used in a metrologically "approved" application, it must be protected
from tampering by use of seals. Optional sealing kits that contain all the required hardware are
available from METTLER TOLEDO. The method used for sealing may vary depending upon the local
requirements.
3.12.1.

Sealing the Panel-Mount Enclosure

The panel-mount enclosure supports both two- and three-screw external sealing for the United
States and Canada. An additional sealing plate, which covers the terminal strips is available for
approved installations in Europe.
3-42
METTLER TOLEDO IND560x Installation Manual
Figure 3-50: Preparing the Capacity Label
Capacity Label
in Place
Figure 3-51: Capacity Label Installed
Cut at dotted line to remove
unused ranges/intervals
64061929
|
11
|
07/2017

Publicidad

Tabla de contenido
loading

Tabla de contenido