3.12.1.1.
United States and Canadian Sealing
To seal the panel-mount enclosure, follow these steps:
1. Ensure that the appropriate approval region has been selected in setup under Scale > Type >
2. Secure the back panel of the IND560x panel mount using the through-hole screws provided in
3. Thread the wire cable and plastic seal through the hole in each new screw.
4. Thread the end of the wire cable through the plastic seal and snap the seal shut. Cut off any
3.12.1.2.
European Sealing
Sealing requirements per OIML standards require additional physical protections of the load cell
connections. A unique kit is available for sealing the panel mount IND560x in approved
installations.
To seal the panel-mount enclosure, follow these steps:
1. Ensure that the appropriate approval region has been selected in setup at Scale > Type >
2. Uninstall any I/O or interface option boards and disconnect their cables from the main board of
3. Remove the 4 Phillips-head screws that normally secure the main board to the permanent
4. Install the 4 brass standoffs included with the sealing kit, onto the permanent standoffs in the
5. Place the horizontal sealing plate on top of these newly installed standoffs. Take care to slide
6. Install any I/O or interface option boards into the IND560x over the top of the horizontal sealing
7. Remove the appropriate adhesive covers from the vertical cover plate, and slide the vertical
8. Secure the back panel of the IND560x using three Phillips-head screws and the two screw
9. Place appropriate paper marks over these sealing sleeve/screw combinations.
64061929
|
11
|
07/2017
Approval and that the Metrology security switch SW1-1 is in the "ON" position.
the sealing kit. The U.S. requires use of only two of the screws; Canada requires use of all
three.
excess wire.
Approval and that the Metrology security switch SW1-1 is in the "ON" position.
the IND560x panel mount. No additional option boards can be installed at this time. The
display cable can remain connected to the main board.
standoffs inside the chassis. Save these screws.
chassis where the 4 Phillips-head screws were just removed.
the front edge of the sealing plate into the correct channel on the panel mount chassis. Secure
the sealing plate into place with the 4 Phillips-head screws that originally held the main board
in place.
In order to get the sealing plate to fit, it may be necessary to loosen the hex screws that hold the panel
mount integrated brackets in place.
plate and reconnect the harnesses to the main board of the IND560x.
cover plate into place.
sealing sleeves. Insert two of the Phillips-head screws through the sealing sleeves and tighten
to the back of the terminal.
METTLER TOLEDO IND560x Installation Manual
3-43