3.3.1.1.
Ferrites for SICSpro Panel Mount Version
For the IND560x SICSpro terminal Panel version, ferrite cores are required to be installed on the
cable wires for COM1 serial communication, discrete I/O, current loop interface and power supply
cable. A single wrap around the outside of the core is only required for the SICSpro scale interface.
Wires for the remaining interfaces can be routed through the core to the connector without the single
wrap around the outside of the ferrite core.
Ferrite cores for the COM1 serial interface (SICSpro panel version only), power supply cable
(SICSpro panel version only), scale interface (all versions in panel and Harsh) and other factory-
configured options are included with the terminal accessory package.
Fiber optic interface cables do not require Ferrite cores.
Figure 3-15 shows the additional ferrites required for the COM1 serial port, discrete I/O, current loop
and power supply in the SICSpro panel version of the IND570 terminal:
3.3.2.
Harsh Enclosure Cable Glands
The IND560x harsh environment terminal is designed to withstand severe washdown and dusty
environments. However, care must be taken when installing cables and/or connectors that enter the
3-10
METTLER TOLEDO IND560x Installation Manual
COM1
Serial Port
Digital I/O
Figure 3-15: Ferrites Required for SICSpro Panel Installation
Table 3-1: Ferrites Required for SICSpro Panel Installation
ONLY THE COMPONENTS SPECIFIED IN THIS MANUAL CAN BE USED IN THIS DEVICE. ALL
EQUIPMENT MUST BE INSTALLED IN ACCORDANCE WITH THE INSTALLATION INSTRUCTIONS
DETAILED IN THIS MANUAL. INCORRECT OR SUBSTITUTE COMPONENTS AND/OR DEVIATION
FROM THESE INSTRUCTIONS CAN IMPAIR THE INTRINSIC SAFETY OF THE TERMINAL AND COULD
RESULT IN BODILY INJURY AND/OR PROPERTY DAMAGE.
Cable
Quantity
COM1 Serial Port
1
Discrete I/O
2
Current Loop
1
SICSpro Scale
1
Power Supply
1
WARNING
Scale
Power
Supply
64061929
|
11
|
07/2017