When connecting to an ACM500, the number of communication channels required depends on the
options that will be used in the ACM500. It may not always be necessary to provide connections for
both circuits. Refer to Table 3-16 to determine if both the COM4 and the COM5 circuit must be
connected. Two fiber optic cables are required for bidirectional communication on each port. If both
COM4 and COM5 are to be used bidirectionally, four lengths of cable must be installed.
The IND560x terminal does not automatically include any fiber optic cables when the Fiber Optic
interface is installed at the factory. Hard clad silica and plastic core fiber optic cables are available
from METTLER TOLEDO.
3-34
METTLER TOLEDO IND560x Installation Manual
Note: When using the IND560x as a remote terminal for a Puma or ID3sTx, the J2 jumper on
the Puma/ID3sTx FO option pcb must be adjusted to send a more intense signal. This is
because the IND560x requires a stronger FO signal to correctly work as a remote display for
the Puma/ID3sTx. To send a stronger signal from the Puma/ID3sTx, position the jumper on the
J2 pins (Figure 3-43) to close the connection between the A, B or C pins until the signal
transmitted by the Puma/ID3sTx is strong enough to be seen by the IND560x remote.
Figure 3-43: J2 Pin Location on Puma/ID3sTx FO Option PCB
Table 3-16: IND560x Interface Port Wiring
Option Used in
ACM500
COM2 Port
COM3 Port
Ethernet
PLC Option
When using fiber optic cable, care must be taken not to bend the cable sharply or it could be
permanently damaged.
COM4
COM5
Required
Required
--
--
--
--