3.3.3.3.
High Precision IDNet Connection
The IND560x terminal IDNet connection supports the new T-Brick Ex T4 type cell used in the K..x
bases and the Point Ex cell. The older types of intrinsically safe modules (known as PIK Ex) which
were also used in the K..x bases, are not compatible with the IND560x terminal.
When using an IDNet version of the IND560x terminal, the cable connection from the base is made
to the J1 and J9 connectors (also identified as the "SD" connection blocks) on the main board as
shown in Figure 3-18, Figure 3-19
The intrinsically safe version of the IDNet base is supplied with an open ended cable for
termination. Table 3-7 shows the wiring color code for the IDNet terminal strip.
3-18
METTLER TOLEDO IND560x Installation Manual
NOTE: When using four-wire load cells, jumpers must be placed between the +Excitation and
+Sense terminals and between the −Excitation and −Sense terminals.
Note for load cell wiring: If an increase in load results in a decrease in weight display, reverse
the signal wires (+SIG and −SIG).
Figure 3-22: IDNet and SICSpro Cable Termination
Table 3-7: IDNet and SICSpro Cable Color Code
Terminal
SD-1
SD -2
SD -3
SD -4
SD -5
SD -6
SD -7
Figure 3-22.
and
Wire Color For TBrick Ex
and MPXI (SICSpro)
Gray
Pink
Yellow
Brown
Green
White
No Connection
Wire Color for Point Ex
No Connection
Pink
Yellow
Brown
Green
White
Gray
64061929
|
11
|
07/2017