Amprobe MultiTest 2000 Manual De Usuario página 154

Ocultar thumbs Ver también para MultiTest 2000:
Tabla de contenido

Publicidad

AMPROBE
Test b):
Connect the instrument
test
leads
to
the
electrodes placed on the
floor
at
a
reciprocal
distance of one meter.
Measurements of floor insulation resistance in medical rooms
The electrodes shall consist of a plate having a surface of 20 cm
approx. 1 Kg (10N), and a humid absorbing paper ( or humid thin cotton cloth) with
the same surface placed between the metal plate and the floor.
The insulation resistance is represented, both for the measurements indicated in "a"
and for the measurements indicated in "b", by the average of 5 or more tests
MULTITEST2000
Test a):
Connect one test lead of
the instrument to the
equalising potential node
and the other to one of
the electrodes placed on
the floor at a distance
higher than one meter
away
from
earthed
objects.
2
, weight equal to

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido