AMPROBE
TEST LEADS
6.1.1.1.
Procedure to reset test leads calibration parameters
To
cancel
calibration
parameters
necessary to perform a
calibration procedure
with a resistance of
test leads higher than
5Ω (for example with
open test leads). When
a cancellation is effected
the screen to the right
is displayed temporarily.
Before each measurement always assure that the
calibration is for the cables in use. During a
continuity test, if the resistance value free of
calibration (that is the resistance value less the
calibration offset value) is negative, the symbolν
is displayed. Probably the calibration resistance
value stored in the instrument memory is not for
the cables in use, therefore a new calibration must
be performed.
LOWΩ
it
is
ν
99.9
>
R+
----Ω
---mA
AUTO
0.11Ω
FUNC
CAL
Message
05.06.01
means
instrument detected a
resistance
Ω
than 5Ω therefore it
will
proceed
R-
Reset procedure.
----Ω
---mA
MULTITEST2000
>99.9Ω:
that
the
higher
with