Amprobe MultiTest 2000 Manual De Usuario página 86

Ocultar thumbs Ver también para MultiTest 2000:
Tabla de contenido

Publicidad

AMPROBE
Symbols
Description
The recording of all the selected parameters will start at 00
START:MAN
seconds after pressing START/STOP (see chapter 10.1).
The recording of all the selected parameters will be interrupted
STOP:MAN
manually by pressing START/STOP (see chapter 10.1).
The recording of all the selected values will be started / interrupted at
START:AUTO
the set dates and times. In order to start the recording the user will
STOP:AUTO
have to press START/STOP to set the instrument in Stand-by mode
until the start date and time previously set (see chapter 10.1).
The value of this parameter determines every how many seconds
the values of all the selected parameters will be memorised
INT. PERIOD
(see chapter 16.8.1). Available choices:
5sec,10sec,30sec,1min,2min,5min,10min,15min,30min,60min.
ON =
HARM REC.
voltage and current harmonics.
OFF = the instrument will not record any voltage or current
ON = the Instrument will record Voltage Anomalies (voltage Sag
ANOM REC.
OFF = the instrument will not record any voltage Sag and Surge
V1
Voltage RMS value
THD, DC, 01...49
Voltage Total Harmonic Distortion, DC Component, 01..49
Harmonics respectively
RMS reference value for Voltage used in Voltage Anomalies
Vref
detection (Voltage Sag and Surge). The Reference is:
(only if ANOM. REC flag
a) Voltage Phase to Neutral for Single Phase and 4 wires three
has been set ON)
b) Voltage Phase to Phase for 3 wires three phase system
High and Low Voltage Percent threshold used in Voltage
Anomalies detection (Voltage Sag and Surge). These parameters
LIM+, LIM-
can be adjusted in range 3% ÷ 30% (step 1%).
(only if ANOM. REC flag
Example: Three Phase System 4 wires.
has been set ON)
Vref = 230, LIM+= 6%, LIM-=10% =>
High Lim = 243.8V, Low Lim = 207.0V
the instrument will record the values of the selected
harmonic selected
and Surge) (see paragraph 16.5)
phase system
MULTITEST2000
Advised settings
15min
V1
THD,01,03,05,07
Single phase: 230V
3 Phase system:
3 wires: 400V
4 wires 230V
+6% / -10%

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido