AMPROBE
8.1.
BASIC SETTING: ANALYZER CONFIG
Place the rotary switch in the ANALYZER position, press the MENU key, using the F1/F2
keys select the ANALYZER CONFIG item and press the ENTER Key.
The following page will be displayed:
This page of setting can be confirmed by pressing the ENTER key or cancelled by
pressing the ESC key.
8.1.1. Type of electrical system under test
The parameter SYSTEM is fixed to "SINGLE" value as Multitest 2000 is able only to
analyse Single Phase system.
8.1.2. How to set the fundamental frequency
Position the cursor on the corresponding word by pressing the multifunction keys F1 and
F2 and select the network frequency between the possible values 50Hz and 60Hz by
pressing the multifunction keys F3 and F4. This parameter is important ONLY if the input
voltage is not sufficient to recognise the value of the frequency (for example, only the
clamps for the current measurement are connected). In this case the instrument generates
an internal synchronism equal to the value of the set frequency.
ANALYZER CONFIG
SYSTEM
:SINGLE
FREQUENCY:50HZ
CURRENT RANGE:1000A
CLAMP TYPE: STD
TV RATIO:0001
PASSWORD:ON
↓
↑
+
-
MULTITEST2000