AMPROBE
Too many records
No Unit selected
ERR: SEQ
ERR: P-
ERR: SEQ & P-
ERR: CONNECTION
Error Vref
ERR: SYNC
Selection Error
Error1 ÷ Error 5
The quantity of recorded data + Smp exceeds the
maximum allowed (35)
The Phase Sequence isn't correct.
The active powers shown on the right side of the message
are negative
The active powers shown on the right side of the message
are negative and the Phase Sequence isn't correct.
The instrument has detected a wrong connection to
Voltage inputs
The user set a Voltage reference not compatible with
voltage at instrument's input.
The System Frequency is out of range
There is a mismatch between the Parameter enabled and
the parameter selected for an AUX recording.
The instrument memory is damaged.
MULTITEST2000
Cancel some recordings after transferring
them to a PC
Check the Phase Sequence connection.
If there isn't a situation of co-generation
check if the clamps are properly connected
If there isn't a situation of co-generation
check if the clamps are properly connected /
check the Phase Sequence connection.
Check the Voltage connections
Check Voltage Reference set in "CONFIG
RECORDER"
Check the System Frequency, check setting
in ANALYZER CONFIG.
Check the parameter enabled in AUX
position and the selected parameter for
recording.
Contact HT Italia assistance