8.5.
"CURRENT" FUNCTION
This function permits to display in real time the RMS value of AC/DC current, the peak, the
Thdl value (see paragraph 16.6), the waveform and the harmonic spectrum of the current.
8.5.1. Symbols
The CURRENT position has three working modes:
METER
WAVE
HARM
These modes will be described in detail in the next paragraphs.
The symbols used are described below:
Symbol
Description
I1
Current RMS value
Ipk1
Peak value of the current
h01 ÷ h49
Harmonic 01 ÷ harmonic 49.
ThdI
Total harmonic distortion factor of the current (see paragraph 16.6).
freq
Network frequency
Tab. 2: Symbols used in the position CURRENT