ECG
ECG is an option and requires a FUJIFILM SonoSite ECG cable.
WARNINGS
Caution
Notes
To use the ECG
1 Connect the ECG cable to the ECG connector on the ultrasound system, mini-dock, or docking system.
The ECG feature turns on automatically.
Note
2 Select ECG on-screen. (ECG may be on another page. It appears only if the ECG cable is connected.)
3 Select options as desired.
Table 2: ECG on-screen controls
Turns on and off the display of the ECG signal.
Show/Hide
Increases or decreases ECG gain. Settings are 0-20.
Gain
Sets the position of the ECG signal.
Position
4
Do not use the FUJIFILM SonoSite ECG to diagnose cardiac arrhythmias to
provide long-term cardiac monitoring.
To avoid electrical interference with aircraft systems, do not use the ECG
cable on aircraft. Such interference may have safety consequences.
Use only accessories recommended by FUJIFILM SonoSite with the system.
Your system can be damaged by connecting an accessory not
recommended by FUJIFILM SonoSite.
Biopsy guidelines are not available when the ECG cable is connected.
The ECG signal may take up to one minute to restabilize after defibrillator use
on a patient.
The ECG signal may take up to one minute to restabilize after defibrillator use
on the patient.
ECG