CHAPTER 9: DIAGNOSTICS
9.1

GENERAL WARNINGS

If any of the anomalies described below should
occur, solve the problem, where allowed, with
respect to all instructions contained in this
manual. If the problem persists, contact the
Authorised After-Sales Centre.
9.2

TROUBLESHOOTING

EN
MALFUNCTION The
induction chamber does not heat up.
CAUSE
1. The instrument is made from a material
which does not conduct heat.
2. The instrument is tarnished (the sensors
detect light colours more easily).
3. The sensors do not "detect" the presence
of the spatula.
4. The power cable is not connected.
5. There is no current to the socket.
6. The appliance is off.
7. Electrical fault.
REMEDY
1. Replace the instrument used.
2. Clean or replace the instrument used.
3. Adjust the sensitivity of the sensors.
4. Connect the power cable.
5. Check the socket/electrical panel in the
building.
6. Turn the appliance on by pressing the
ON/OFF button.
7. Contact an authorised service centre.
MALFUNCTION Red LED flashing rapidly.
CAUSE
1. An object has been left for too long in
the induction chamber.
2. The sensor is disorientated.
3. Electrical fault.
REMEDY
1. Remove the object from the induction
chamber.
2. Turn the tool heater off, then on again.
If the problem continues, adjust the
sensitivity of the sensors.
3. Contact an authorised service centre.
MALFUNCTION The cooling fan makes a different
noise from usual.
CAUSE
1. Appliance overheated.
2. Faulty fan.
REMEDY
1. Turn the appliance off and allow it to
cool down.
2. Contact an authorised service centre.
22
The reference figures can be found on pages 3 and 4 of this manual.
instrument
placed
in
CHAPTER 10: ATTACHMENTS

10.1 EC DECLARATION OF CONFORMITY

The Manufacturer:
Via Bovazecchino, 100
45021 Badia Polesine • RO • Italy
Tel. +39 0425 597611 • Fax +39 0425 53596
Person authorised to compile the technical file:
the
Via Bovazecchino, 100
45021 Badia Polesine • RO • Italy
Tel. +39 0425 597611 • Fax +39 0425 53596
declares under its sole responsibility that the product:
is in conformity with Directive 2011/65 EU of the European
Parliament and of the Council of 8 June 2011 on the
restriction of the use of certain hazardous substances in
electrical and electronic equipment.
It is also in conformity with the laws implementing the
following directives:
· EC Directive 2014/30/UE (EMC Directive) and subsequent
amendments.
· EC Directive 2014/35/UE (Low Voltage Directive) and
subsequent amendments.
According to the following harmonised standards:
EN 61326-1:2013
EN 61000-3-2:2007+A1:2011
EN 61000-3-3:2014
EN 61010-1:2010
Please note that the manufacturer's liability for damages
caused by a defective product expires 10 years from the
date on which the manufacturer released such product
on the market and that such liability is transferred to the
user in accordance with Directive 85/374/EEC, subsequent
amendments and the laws implementing it.
Badia Polesine, 20/04/2016
Zhermack S.p.A.
Zhermack S.p.A.
cyclope
Paolo Ambrosini
General Manager
cyclope

Publicidad

Tabla de contenido
loading

Tabla de contenido