ASTUTE140
Meter
®
If all systems pass, the Log In screen will automatically appear, indicating that the
ASTUTE140
Figure 15.
Log In screen
If the self test fails, an error message will be displayed. Press the right soft key to select
OK and close the error message. Restart the ASTUTE140
back on again (to power off the meter, toggle the power switch upwards and it should
immediately shut down). If after restarting the ASTUTE140
again, contact Astute Technical Support (See "Contacting Astute Medical, Inc." on p. 7
for contact information).
ASTUTE140
Test cartridge will provide equivalent results if read in different meters. The End User is
not required to calibrate the meter.
Supervisor Instructions:
Configuration and Settings
Adding the First Supervisor User
The first Supervisor user must be added to the ASTUTE140
settings may be set or changed. The ASTUTE140
a default Supervisor log in ID (1234) that will allow the first Supervisor user to log in
and create a user profile. Once the new profile is created, the user should log out of the
system and log back in with the new user information before performing any additional
functions. After this process is completed, the default log in (1234) will need to be
manually deleted.
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
Meter is ready for use.
®
Meters are factory-calibrated to ensure that the same NEPHROCHECK
®
Meter by turning it off and
®
Meter the self test fails
®
Meter before any default
®
®
Meter comes from the factory with
Installation
®
23