See the assay-specific Package Insert for storage instructions when the liquid control
kit is not in use. Additional liquid controls may be ordered from Astute Medical, Inc. at
www.astutemedical.com or by contacting your Astute Medical, Inc. sales representative.
For each NEPHROCHECK
must be carried out twice: once for the high liquid control and once for the low liquid
control. To register each control, perform the following steps:
1. Locate the NEPHROCHECK
lot to be registered (i.e. the NEPHROCHECK
registering the high liquid control, or the NEPHROCHECK
RFID Card if registering the low liquid control).
2. Press the
lot immediately after successful log in, the Main Menu will automatically be
displayed).
3. Use the navigation keys to highlight the Operator Menu icon.
Figure 60.
Main Menu screen with
Operator Menu icon
highlighted
4. Press the right soft key to display the Operator Menu.
5. When the Operator Menu is displayed, Manage Lots is highlighted. Press the right
soft key to display the Manage Lots screen.
6. When the Manage Lots screen is displayed, Manage Test Lots is highlighted. Use
the
the Registered LQC Lots screen.
7. On the Registered LQC Lots screen, press the right soft key to display the Options
pop-up menu.
8. When the Options pop-up menu is displayed, Print List is highlighted. Use the
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
Liquid Control Kit, the liquid control registration process
®
key to display the Main Menu (if registering the liquid control
key to highlight Manage LQC Lots and press the right soft key to display
key to highlight Register Lot and press the right soft key.
ASTUTE140
Liquid Control RFID Card for the liquid control
®
High Liquid Control RFID Card if
®
Meter Operation
®
®
Low Liquid Control
55