4. If available, a Sample ID may be entered; otherwise, this field may be left blank. To
enter a Sample ID, use the
ID manually using the numeric keypad or an external keyboard (if connected)
or, if applicable, by scanning the barcode on the patient sample with an optional
barcode reader (if connected). NOTE: an external keyboard and barcode reader
are optional devices that are not included with the ASTUTE140
5. After confirming that the correct Patient ID and/or Sample ID have been entered,
press the right soft key to select Run Patient.
6. After selecting Run Patient, the ASTUTE140
open.
7. Prepare the NEPHROCHECK
to the NEPHROCHECK
Test cartridge must be run immediately after inoculation with the patient sample,
and the next cartridge should not be prepared until testing on the first has
completed. The ASTUTE140
only one test can be run at a time. Inoculating additional samples before testing
has completed may lead to inaccurate results.
8. When prompted, place the prepared NEPHROCHECK
recess inside the ASTUTE140
on the NEPHROCHECK
drawer until it clicks.
Figure 67.
Insert Cartridge screen
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
key to highlight Sample ID and enter the Sample
®
Meter drawer will automatically
Test cartridge with the patient specimen according
®
Text Kit Package Insert. NOTE: Each NEPHROCHECK
®
®
Meter times the incubation period for the test, and
Meter drawer with the Astute Medical logo marked
®
Test cartridge pointed toward the meter. Gently close the
®
ASTUTE140
Meter Operation
®
Meter kit.
®
Test cartridge into the
®
®
63