Error Messages - Astute Medical ASTUTE140 METER Manual Del Usuario

Ocultar thumbs Ver también para ASTUTE140 METER:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Maintenance and Troubleshooting

Error Messages

Type
Screen Display
*Error codes may begin with 001, 002, 003 or 004.
76
Error Code*
What it Means
Start-up/
00X.060.128
Self-Test
Error.
Invalid User
00X.061.086
Error. User is
not stored or
does not have
permission to
perform the
function.
Invalid User
00X.061.085
Error. User is
expired.
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
ASTUTE140
Action to Take
Power the
ASTUTE140
®
Meter
off and then on
again. If the self-test
fails a second time,
discontinue use
and call Astute
Technical Support
(See "Contacting
Astute Medical, Inc."
on p. 7 for contact
information).
Carefully re-enter
a valid user ID and
password.
Check with
Supervisor regarding
permissions granted.
Supervisor to create
a new user.
If the error persists,
call Astute Technical
Support (See
Contacting Astute
Medical, Inc." on
p. 7 for contact
information).
Carefully re-enter
a valid user ID and
password.
Create a new user.
If the error persists,
call Astute Technical
Support (See
Contacting Astute
Medical, Inc." on
p. 7 for contact
information).
Meter
®

Publicidad

Capítulos

Tabla de contenido
loading

Este manual también es adecuado para:

500017