9. After closing the drawer, a temporary screen will appear indicating that the
ASTUTE140
is registered and whether high and low LQC measurements have passed for the
test lot.
a. If the lot has not been registered, an error message will appear. Press the
right soft key to select OK and close the error message. Register the lot
(See "Test Lot Registration" on p. 52) before running the test again.
b. Unless QC Settings are changed by the Supervisor, if the lot has been
registered, but high and low LQC measurements have not passed for the
NEPHROCHECK
indicating that passed LQC measurements are required will be displayed.
If high and low LQC measurements are required, follow these steps:
i. Press the right soft key to select OK and close the error message.
ii. Discard the NEPHROCHECK
64
Meter Operation
Meter is reading the test lot information to determine if the test lot
®
Test Kit lot, the patient test will not be run and a message
®
sample.
Test cartridge prepared with the patient
®
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
ASTUTE140
Meter
®
Figure 68.
Temporary "Reading Test
Lot Info. Please Wait!"
screen
Figure 69.
"Cassette Lot Locked"
error message