Error Messages
Type
*Error codes may begin with 001, 002, 003 or 004.
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
(Continued)
Screen Display
Maintenance and Troubleshooting
Error Code*
What it Means
Measurement
00X.061.060
Locked. A
passing EQC
measurement is
required.
00X.050.005
No or
damaged RFID
Tag.
00X.061.073
Invalid RFID
00X.061.074
Tag.
00X.061.075
Action to Take
Perform an EQC
measurement
(See "ASTUTE140
®
Electronic Quality
Control (EQC) on
p. 50).
If it fails repeatedly,
call Astute Technical
Support. (See
"Contacting Astute
Medical, Inc." on
p. 7 for contact
information).
Position the RFID
card over the Meter
keyboard and retry.
If the problem
persists, call Astute
Technical Support.
(See "Contacting
Astute Medical, Inc."
on p. 7 for contact
information).
Locate and position
the correct RFID
card over the Meter
keyboard and retry.
If the problem
persists, call Astute
Technical Support.
(See "Contacting
Astute Medical, Inc."
on p. 7 for contact
information).
83