ASTUTE140
®
b. When the Options pop-up menu is displayed, Print is highlighted. To print
the LQC results, press the right soft key.
c. To hide the Options pop-up menu, use the
Options and press the right soft key.
d. To return to the Main Menu, press the left soft key.
13. To run the next liquid control (high or low) after viewing or printing the LQC
results, press the right soft key to display the Options pop-up menu. Use the
key to highlight Next Control and press the right soft key. Repeat steps 4–12.
Calibration Verification
A NEPHROCHECK
States only) is available to assist the laboratory in monitoring test performance. Good
Laboratory Practice suggests that the NEPHROCHECK
should be tested at least once every 6 months, after ASTUTE140
in accordance with the laboratory standard procedures.
The NEPHROCHECK
procedures for preparing and testing patient samples with the NEPHROCHECK
and ASTUTE140
To test calibration verification materials, perform the following steps for each of the
vials:
1. Press the
immediately after successful log in, the Main Menu will automatically be
displayed).
2. When the Main Menu is displayed, use the navigation keys to highlight the Run
Patient icon. Press the right soft key to select Run Patient.
60
Meter Operation
Calibration Verification (Cal Vers) Kit (Available in the United
®
®
Calibration Verification vials should be prepared following the
Meter.
®
key to display the Main Menu (if running calibration verification
key to highlight Hide
(Available in the United States only)
Calibration Verification Kit
®
®
Meter servicing, and
Test
®
Figure 65.
Main Menu screen
with Run Patient icon
highlighted
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
ASTUTE140