Astute Medical ASTUTE140 METER Manual Del Usuario página 82

Ocultar thumbs Ver también para ASTUTE140 METER:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Maintenance and Troubleshooting
Error Messages
Type
Screen Display
*Error codes may begin with 001, 002, 003 or 004.
82
(Continued)
Error Code*
00X.061.073
00X.061.031
00X.061.025
00X.061.083
What it Means
Action to Take
Test Type
Passing positive
Mismatch.
and negative LQC
The LQC and the
measurements
Cartridge have
for the test lot is
different test
required.
types.
Retest using a
Cartridge for which
LQC has been
performed and
passed.
Register and perform
testing with a new
LQC lot.
Unknown
Kits and devices
Lot Error. The
must be registered
required Lot is
prior to running
not installed.
(See "RFID Cards
for Device and Lot
registration" on p. 48
for instructions).
Lot Already
The Lot does
Registered.
not need to be
registered. Proceed
with testing.
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07
ASTUTE140
Meter
®

Publicidad

Capítulos

Tabla de contenido
loading

Productos relacionados para Astute Medical ASTUTE140 METER

Este manual también es adecuado para:

500017

Tabla de contenido