4. A screen showing all liquid control lots previously registered and their expiration
dates will be displayed. Controls displayed as "Control +" correspond to high liquid
controls and those displayed as "Control –" correspond to low liquid controls.
5. Use the
press the right soft key. If the liquid control lot is not listed, register the liquid
control lot to be tested by following the steps outlined in "Liquid Control Lot
Registration" on p. 54. After the lot is registered, repeat steps 1–6.
6. After selecting the liquid control lot to be tested, the ASTUTE140
will automatically open.
7. Prepare the NEPHROCHECK
as instructed in the assay-specifi c NEPHROCHECK
Insert. NOTE: The test lot must be registered prior to running LQC (See "Test Lot
Registration" on p. 52 for instructions).
58
Meter Operation
keys to highlight the correct liquid control lot being tested and
Test cartridge with the liquid control (high or low)
®
Figure 63.
Run LQC — Select Lot
screen
Meter drawer
®
Liquid Control Kit Package
®