ASTUTE140
®
8. When prompted, hold the ASTUTE140
Card for the ASTUTE140
register the ASTUTE140
right soft key.
9. If registered correctly, a screen indicating that the ASTUTE140
number was successfully read from the ASTUTE140
Press the right soft key to select Accept. The ASTUTE140
just registered should now appear in the list of registered ASTUTE140
Devices.
10. If registered incorrectly, an error message will appear. Press the right soft key to
select OK and close the error message. Repeat steps 8–9. If registered incorrectly
a second time, contact Astute Technical Support (See "Contacting Astute Medical,
Inc." on p. 7 for contact information).
11. After use, return the ASTUTE140
the ASTUTE140
ASTUTE140
The EQC procedure verifies the calibration of the ASTUTE140
positioning system, optical system and other internal systems of the meter to confirm
that the meter is functioning properly. It is recommended that the EQC procedure be
run daily. The required frequency for running EQC may be set by a Supervisor in the
general ASTUTE140
instructions).
Prior to running the first EQC procedure, the ASTUTE140
registered (See "ASTUTE140
50
Meter Operation
EQC Device next to or against the numeric keypad to
®
®
EQC Device information and select OK by pressing the
EQC Device in its original packaging.
®
Electronic Quality Control (EQC)
®
Meter default settings (See "Quality Control Settings" on p. 31 for
®
EQC Device Registration" on p. 48 for details).
®
Electronic Quality Control (EQC) RFID
®
EQC RFID Card will appear.
®
®
EQC RFID Card to its sleeve and store it with
®
EQC Device must be
®
Figure 55.
Register Electronic QC
Device screen
EQC Device
®
®
EQC Device that was
®
EQC
Meter, including the
©Astute Medical, Inc. 2020 PN 300153MULTI Rev A 2020/01/07