7.8
Automatic cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must fundamentally have a tested
efficacy (such as FDA approval or CE label pursuant to DIN EN ISO 15883).
7.8.1
Manual pre-cleaning with a brush
Phase
Step
I
Rinsing
II
Brushes
D–W:
Drinking water
RT:
Room temperature
►
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
Phase I
►
Mobilize non-rigid components during cleaning.
►
Thoroughly clean the product under running water.
Phase II
►
Mobilize non-rigid components during cleaning.
►
Brush difficult to access surfaces with a suitable plastic cleaning brush
for at least 1 min.
7.8.2
Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection
V
Drying
DW:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
*Recommended: BBraun Helimatic Cleaner alkaline
►
Insert the product in its correct position in the ECCOS holder.
►
Screw the rinsing adapter (GA672211) on the product in the mount 14.
►
Then connect the interior rinse device and connect to the rinse connec-
tor of the cleaning/disinfection device/rinse cart.
►
Attach the Kirschner wire protection sleeve 15 to a rinse hose.
►
After automatic cleaning/disinfection, check visible surfaces for resi-
dues and repeat the cleaning/disinfection process as needed.
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold)
–
–
RT (cold)
–
–
T
t
[°C/°F]
[min]
<25/77
3
55/131
10
>10/50
1
90/194
5
max.
min. 10min
120/248
Note
The cleaning and disinfection machine used for processing must be serviced
and checked at regular intervals.
Water
Chemistry/Note
quality
until visually clean
D–W
D–W
until visually clean
►
After manual preparation, check visible surfaces for residue and repeat
the pre-cleaning process as needed.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available in Aesculap Extranet at
https://extranet.bbraun.com).
Water qual-
Chemical
ity
D–W
–
■
FD–W
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
■
0.5 % working solution
– pH ~ 11*
FD-W
–
FD-W
–
–
–
en
9