Fault
Cause
Oil spray adapter cannot
Oil spray adapter not compat-
be attached
ible
Speed control knob can-
Product in locked OFF position Safety catch is in OFF position
not be actuated
Speed control knob
jammed/broken
Knob for counter-clock-
Counterclockwise rotation
wise rotation cannot be
knob jammed/broken
actuated
10. Technical Service
DANGER
Danger to life of patients and users in case of malfunctions and/or
failure of protective measures!
Do not perform any servicing or maintenance work under any cir-
cumstances while the product is being used on a patient.
CAUTION
Modifications of medical devices may result in a loss of potential
guarantee/warranty claims and forfeiture of applicable licenses.
Do not modify the product.
Contact national B. Braun/Aesculap representative for service and
repair.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Other service addresses can be obtained from the address indicated above.
12
Detection
Oil spray adapter cannot be attached
Speed control knob not working
Left rotation knob not working
11. Accessories/Spare parts
Item no.
GA672207
GA672211
GA675
GA676
GA678
GA679
GB243R
GB244R
GB600
GB600840
TA011944
GB495R
GB487R
GB488R
GB489R
TA014538
TA014539
TA014540
TA014541
Remedy
Use the compatible oil spray adapter for GA330.
Turn the safety catch into the ON position.
Have the manufacturer repair the product.
Have the manufacturer repair the product.
Name
Kirschner wire protection sleeve
Rinsing adapter
Lid
NiMH battery long
Sterile funnel
Battery removal device
ECCOS sieve basket with holder for Acculan 4
STERILIT Power Systems oil spray
Oil spray adapter
Cleaning brush
ECCOS holder for machine
ECCOS holder for lid
ECCOS holder for sterile funnel
ECCOS holder for battery removal device
Instructions for use of the drill and milling machine
GA330 (A4 for ring binder)
Instructions for use of the drill and reamer GA330
(flyer)
Instructions for use for attachments (A4 for ring
folders)
Instructions for use for attachments (flyer)