Unidrive ® Troubleshooting Table - Storz UNIDRIVE S III ENT Manual De Uso

Ocultar thumbs Ver también para UNIDRIVE S III ENT:
Tabla de contenido

Publicidad

79
Technische Beschreibung
10. 2 UNIDRIVE
troubleshooting
®
table
Error message
E01
MEMORY ERROR
E02
MOTORBOARD DEFECTIVE
E03
HARDWARE CONFIGURATION ERROR
E04
SOFTWARE CONFIGURATION ERROR
E05
-
E06
-
E07
-
E08
-
E09
MOTOR DEFECTIVE M1
E10
MOTOR DEFECTIVE M2
E11
NO FOOTSWITCH
E12
NO MOTOR
E13
TOO HIGH TEMPERATURE MOTOR
M1
E14
-
E15
TOO HIGH TEMPERATURE MOTOR
M2
E16
IRRIGATION PUMP NOT CLOSED
E17
MOTOR NOT SUPPORTED M1
E18
MOTOR NOT SUPPORTED M2
E19
MOTOR STOPPED M1
E20
MOTOR STOPPED M2
E11
NO SCB MODULE
Technical description
Cause
Parameters could not be saved successfully
Motor control overloaded
Hardware configuration is incorrect
Software configuration is incorrect
Motor on channel 1 is defective
Motor on channel 2 is defective
Footswitch not connected correctly
No motor or handpiece connected
Motor or handpiece on channel 1 overheated
Motor or handpiece on channel 2 overheated
Pump lever not connected correctly
The motor or handpiece connected on channel 1
is not supported.
The motor or handpiece connected on channel 2
is not supported.
Motor or handpiece on channel 1 blocked in
operation
Motor or handpiece on channel 2 blocked in
operation
Device is not connected to an active SCB bus.
Descripciones técnicas
Troubleshooting
Turn device on/off. Load default values. Or send
device in for repair.
Turn device on/off. Or send device in for repair.
Send device in for repair.
Send device in for repair.
Release footswitch and restart.
If the error persists, a different handpiece must be
used.
See E09
Connect footswitch.
If the error persists, check whether the error is on the
footswitch or device. Send the defective component
in for repair.
Allow the connected handpiece to cool down or
operate an additional handpiece on the 2
nd
See E13
Set pump lever vertically. If the error persists, the
device must be sent in for repair.
Only use ones approved by KARL STORZ. An
update by Service may be necessary.
See E17
Release footswitch and restart. If the error persists,
clean the handpiece or exchange the attachment.
See E19
Insert SCB cable and activate SCB bus.
channel.