7.8
Topping up test oil
The test oil must be topped up if the level in the test
oil tank drops below the mark on the oil level indica-
tor (Fig. 1, item 4).
1. Remove the screwed drain plug from the test oil
filler port (Fig. 3, item 9).
!
If the test oil is poured in too quickly, it may run
out at the filler port. Immediately clean up test oil
spills with a cloth.
!
Only fill the EPS 118 with test oil according to
ISO 4113. Never use diesel fuel.
!
The test oil level must be both above the mark of
the oil level indicator (Fig. 1, item 4) and within
the sight glass. If the test oil level cannot be seen
in the within the sight glass, too much test oil has
been filled, which will distort the measured values.
2. Pour test oil (ISO 4113) through the funnel very
slowly and carefully into the test oil filler port
(Fig. 3, item 9). Add test oil until it is above the
mark on the oil level indicator (Fig. 1, item 4).
3. Screw the knurled nut clockwise into the test oil
filler port (Fig. 3, item 9).
"
The EPS 118 is ready for operation.
Robert Bosch GmbH
Operation | EPS 118 | 81
7.9
Faults
!
Do not make any settings in the operating system
or delete data. Do not execute the functions "Re-
turn to factory settings" or "Autom. restore" either.
Otherwise, the display and HMI must be replaced
by customer service at a charge and EPS 118 reca-
librated.
i
Follow the instructions on the screen with regard
to faults in the EPS 118 software. Customer service
must be notified if the problem cannot be rectified.
Fault
LCD remains blank following
switch-on.
Measured values implausible
The "Error" symbol (
) is dis-
played.
Quantity too low during test
of CRIN
Severe quantity fluctuations
during the test of a CRIN with
one-piece armature.
en
Remedial action
Check electrical connection
(if power is being supplied
properly, the LED in the injec-
tion chamber must light).
Check the amount of test oil.
Bleed the system (see sec.
8.2.8).
Change the strainer in all
hose assemblies (see sec.
8.2.9).
Compressed air supply too
low or fluctuates excessively.
Check the compressed air
supply (for prerequisites,
see sec. 5.1.1).
Check whether the "CRIN ad-
apter with check valve" is
used for the test (see sec.
5.4.9). If not, use the "CRIN
adapter with check valve" and
repeat the test.
Check whether the "CRIN ad-
apter with check valve" is
used for the test (see sec.
5.4.9). If yes, use the "CRIN
adapter with check valve" and
repeat the test.
|
1 689 989 300
2020-03-24