Vercise™ Neural Navigator 3 Programming Manual

Starting a Programming Session

Starting Vercise Neural Navigator

1. Power ON the CP and log in as ClinicUser.
2. Select the Vercise Launcher icon
3. Select
to start Vercise Neural Navigator.
Note: If Brainlab Elements is present on the CP, Vercise Neural Navigator may be launched from within Elements
Note: Multiple softwares should not be run simultaneously on the same CP (except when launching Vercise Neural Navigator from within Elements).
Note: Vercise Neural Navigator can also be launched in Demo Mode using the Vercise Launcher. Demo Mode is used for demonstration purposes only
(Figure 2).
4. Upon starting Vercise Neural Navigator, the screen will show the Connect Tab and the software will automatically attempt to connect to a Stimulator
(Figure 3).
(a). If no Stimulator is found, move the Wand closer to the Stimulator that you are trying to connect to and select the Rescan button.
Note: The CP cannot connect to the Vercise Gevia
to connect. For instructions on exiting MRI Mode, refer to the Remote Control Manual as listed in your DBS Reference Guide.
#
Feature
Description
1
Connect Tab
Displays the connection status between the CP, Wand, and Stimulator.
Configure Leads and edit patient profile.
2
Configure Tab
Note: If Brainlab Elements is present on the CP, some patient data may be imported from Elements.
3
Program Tab
Adjust the Stimulator program settings.
4
Data Tab
Generate, print and export reports, and export or delete selected patient(s) data.
5
Tools Tab
Delete patient data.
6
End Session Tab
Disconnect from the Stimulator or exit the application.
7
Battery Indicator
Displays the battery status of the CP.
Vercise™ Neural Navigator 3 Programming Manual
92348958-06 2 of 477
on the desktop.
Figure 2. Launch Screen with DEMO Option Mode
Stimulator when the Stimulator is in MRI Mode. Exit MRI Mode using the Remote Control and rescan
TM
Figure 3. Connect Tab
Table 1: Vercise Neural Navigator Tab Descriptions
.