Descargar Imprimir esta página

Boston Scientific Vercise Instrucciones De Uso página 12

Ocultar thumbs Ver también para Vercise:

Publicidad

Vercise™ Neural Navigator 3 Programming Manual
2. Select a Contact to assign it as a 100% cathode.
3. Use the
and
buttons to incrementally steer the stimulation focus along the length of the Lead. The amount of cathodic current will shift in 10%
increments.
Note: You may also adjust directly from one Contact (Level) to another. The Amplitude for the selected Area will drop to 0 mA when another Contact is
selected, but not when steering in 10% increments.
The DB‑2202 Directional Lead has a total of eight Contacts per Lead, labeled 1 through 8 on each Lead. Contacts 1 and 8 are the distal and proximal Contacts,
respectively, while Contacts 2 through 7 are the small directional Contacts (segments) for each Lead.
To Steer along the DB2202‑Directional Lead:
1. Select Steering Mode.
2. Select a Contact to assign it as a cathode. You may create an equal spread of current across a Level of Contacts ("ring mode") by selecting anywhere
within that Level, then selecting the center button. To assign a single directional segment as the cathode, select anywhere within that Level, then select the
corresponding button (Figure 8).
3. Use the
and
buttons to incrementally steer the stimulation focus along the length of the Lead.
4. Select one of five preset directions
four orthogonal directions or put the stimulation field into "ring mode."
The following steps can be used to refine the applied directional preset or selected directional segment.
5. Use the
and
buttons to rotate and steer the stimulation focus circumferentially around the Lead.
6. Use the
and
buttons to radially spread or shrink the focus of the stimulation field.
7. To choose another starting point or to steer on another Contact, select another Contact. To select a segmented Directional Contact, select the Level, then
select one of the three labeled segmented Contacts around the circumference of the center button on the Directional Contact Selector.
Note: The total Amplitude for the selected Area will drop to 0 mA when another Contact is selected.
Note: Stimulation using multiple independent current control and the DB‑2202 Directional Lead is referred to as Cartesia 3D.
Custom Mode
Custom Mode allows you to assign a percentage of anodic or cathodic current to individual Contacts and the Stimulator Case.
To program the DB2201‑Standard Lead and the DB2202‑Directional Lead in Custom Mode:
1. Select Custom Mode.
2. Select the Stimulator Case or Contact that you want to adjust. If it was selected, one tap will assign it as an anode (+). Another tap will reassign it as a
cathode (‑). Another tap will reassign it as OFF (blank). Tapping on a Contact will first select it without changing the polarity.
Note: Changing the Contact polarities will reset the Amplitude to zero.
3. Select the + and – buttons for the Contact to adjust the percentage of anodic or cathodic current assigned to the selected Contact.
Note: When using the External Trial Stimulator (ETS), monopolar configurations are not possible since the ETS "case" cannot be assigned as a cathode or
anode.
Note: When using the ETS, Clinical Effects data is recorded but not plotted on the CEM.
Vercise™ Neural Navigator 3 Programming Manual
92348958-06 8 of 477
Figure 8. Directional Contact Selector
for the stimulation field. The directional presets will steer the fully focused stimulation field in one of

Publicidad

loading