Digital Communication; Hart Protocol - ABB CoriolisMaster FCB300 Serie Instrucciónes De Puesta En Marcha

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 165
5.5

Digital communication

5.5.1

HART protocol

The device is registered with the HART Communication
Foundation.
4 - 20 mA
Rb = 250 Ohm
BELL 202
Modem
PLS / PC
Fig. 20: Communication via HART protocol
HART protocol
Configuration
— Directly on the device
— Via software DSV401 + HART-DTM
Transmission
FSK modulation on voltage output 4 ... 20 mA
according to Bell 202 standard
Baud rate
1200 baud
Display
Logic 1: 1200 Hz
Logic 0: 2200 Hz
Maximum signal
1.2 mAss
amplitude
Load at current output
250 ... 560 Ω
(in hazardous area: maximum 300 Ω)
Cable
Design
Two-wire cable AWG 24, twisted
Maximum length
1500 m (4921 ft)
See the interface description for detailed information.
HART
G10052
System integration:
Communication (configuration, parameterization) can be
performed with the DTM (Device Type Manager) available for
the device and the corresponding framework applications as
per FDT 0.98 or 1.2 (DSV401 R2).
Other tool/system integrations (e.g., Emerson AMS/Siemens
PCS7) are available on request.
The necessary DTMs can also be downloaded from
www.abb.com/flow.
CoriolisMaster FCB330, FCB350 | CI/FCB300-X1 Rev. C EN - 23

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Coriolismaster fcb330Coriolismaster fcb350

Tabla de contenido