ELECTROMAGNETIC COMPATIBILITY

Warnings
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify
that they are operating normally.
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
Portable RF communications equipment (including peripherals such as antenna cables and external antennas)
should be used no closer than
manufacturer. Otherwise, degradation of the performance of this equipment could result.
Precaution
The device may be susceptible to electromagnetic interference from portable and mobile RF communications
devices such as mobile (cellular) telephones or other equipment, even if that other equipment complies with
CISPR EMISSION requirements.
The PROPRIO FOOT
®
possible, or where exposure to highly electrical and/or magnetic fields can occur (e.g. electrical transformers, high-
power radio/TV transmitters, RF surgical equipment, CT and MRI scanners).
Electromagnetic Emissions Compliance Information
Emissions Test
Conducted and Radiated RF Emissions
Harmonics Emissions
Voltage Fluctuations / Flicker Emissions
cm ( inches) to any part of the device, including cables specified by the
is suitable for use in any environment except where immersion in water or any other fluid is
Standard
CISPR
IEC
- -
IEC
- -
Compliance Level
Group – Class B
Class A
Complies