PERFORMANCE SPECIFICATIONS: The table below shows Arms (Average Root Mean Square) values measured using the E1 Ear Sensor with Masimo SET Oximetry
Technology in a clinical study.
SaO2 versus error (SpO2 – SaO2) with linear regression fit and upper 95% and lower 95% limits of agreement
E1 Sensor
SpO
Arms
2
90-100%
2.04 (2,04) %
80-90%
2.06 (2,06) %
70-80%
2.52 (2,52) %
E1 Sensor