TRACHEOTEC
TRACHEOSTOMY TUBE
I. FOREWORD
These instructions for use are valid for Fahl
inform the physician, nursing staff and patient/user in order to ensure technically correct handling of
Fahl
tracheostomy tubes.
®
Please read the instructions for use carefully before using the product for the first time!
Keep them in an easily accessible place for future reference.
Please keep the package for as long as you use the tracheostomy tube. It contains important infor-
mation on the product!
II. INTENDED USE
Fahl
tracheostomy tubes are intended to stabilise the tracheostoma following laryngectomy or tra-
®
cheostomy.
The tracheostomy tube is designed to keep the tracheostoma open.
Tracheostomy tubes with cuff are always indicated after tracheostomy has been performed for any
cause if the cannula needs to be sealed against the wall of the trachea.
When used for the first time the products must only be selected, used and inserted by clinicians or
medical professionals trained in their use.
Users must be trained by medical professionals on how to use the Fahl
safely and appropriately.
Fahl
LINGO tube variants are intended for tracheotomised patients who retain their larynx or for laryn-
®
gectomees who use a shunt valve (voice prosthesis users).
III. WARNINGS
Patients must be trained by medical professionals on how to use the Fahl
safely.
Fahl
tracheostomy tubes must never be blocked, e.g. by mucus or encrustation. Blockage can
®
cause asphyxiation!
Mucus in the trachea can be aspirated (removed by suction) through the tracheostomy tube
with a tracheal suction catheter.
Damaged tracheostomy tubes must not be used and must be discarded immediately. Use of a
damaged tube may result in airway compromise.
When the tracheostomy tubes are inserted or removed, this can lead to irritations, coughing or
slight bleeding. If bleeding persists, consult your doctor immediately!
Tracheostomy tubes are single­patient products and intended for use by one patient only.
Do not re­use with another patient. Do not reprocess for re­use by another patient.
Do not use tracheostomy tubes during laser treatment (laser therapy) or with electrosurgical
devices! Contact with a laser beam may damage the tube.
Unintentional Inhalation of a tube that was not sized correctly may need to be removed by a
doctor/physician.
In case of an unstable tracheostoma, the airways must always be secured before removing the
tracheostomy tube and a replacement tube must be kept ready for insertion. The replacement
tube must be inserted immediately, even before cleaning and disinfecting the changed tube.
CAUTION!
Tracheostomy tubes with speech function are only recommended for tracheotomised patients
with normal secretion and normal mucous tissue.
Tracheostomy tubes that contain metal parts must under no circumstances be used during a
radiation therapy (radiotherapy), as this can lead to serious skin damage! If you need to wear a
tracheostomy tube during a radiation therapy, then only use plastic tracheostomy tubes with-
out any metal parts. In the case of plastic tracheostomy tubes with silver speaking valve, the
valve can for instance be removed completely from the tube, including the safety chain, by
removing the inner cannula from the outer cannula before starting a radiation therapy.
CAUTION!
If there is strong secretion, a tendency to form granulation tissue, if the patient is undergoing
radiation therapy or if scabs have formed, a perforated cannula version is only recommend-
ed under regular surveillance by a doctor and provided that shorter replacement intervals are
observed (as a rule once a week), because the perforations in the outer tube can increase the
formation of granulation tissue.
IV. COMPLICATIONS
The following complications could emerge when using this device:
Contamination of the stoma may make it necessary to remove the tube. Contaminations can further-
more cause infections that necessitate the use of antibiotics.