10.
SYSTEM SETUP
CONTROLLER SET UP INSTRUCTIONS
1.
Place the Tricera Controller on a stable flat work surface.
IMPORTANT: Prior to use verify that the Controller and footswitch are decontaminated and clean and that
the shaver handpiece is clean and sterilized.
2.
Connect the Tricera Controller power cord (Fig.9) to the power entry module (Fig. 2, Item 10).
a.
Optionally, set up the ZEOS system by connecting the ZEOS console to the rear of the Tricera
Controller via the USB port (Fig. 2, item 12). See Annexe B for ZEOS interface details.
3.
Connect the Tricera footswitch cable to the footswitch receptacle (Fig. 1, Item 1) on the left-hand side of the
front panel of the Controller (Fig. 10).
4.
Turn on the Tricera Controller using the power switch (Fig. 1, Item 7) on the front of the Controller.
a.
Optionally, open the profile screen and select a user profile (See Annexe H for user profile details).
5.
The Zimmer Biomet logo and self-test progress will appear on the screen, followed by the images shown in
Figure 11 and 12.
6. Open the FMA Cassette packaging and dump the contents onto the sterile table.
When using Veriflow (R-10003): Within the sterile field, retain the patient ends of the Inflow and suction connectors
(red and blue connectors). Pass the FMA cassette out of the sterile field with the saline spikes and waste outflow
tube.
When using Exoflow (R-10017): Within the sterile field, retain the patient end of the suction connector (red
connector). Pass the FMA cassette out of the sterile field with the waste outflow tube.
Connect the FMA cassette to the controller pumps by placing the cassette in place as shown. The controller will
detect the presence of the FMA cassette and will provide an audible tone when the cassette is in the correct
position.
DCO-1394
Effective Date: 12-JAN-2022
Figure 9
Figure 10
Figure 11
Figure 12
7. Press the lock button and the FMA cassette will automatically lock into place.
NOTE: for Exoflow Outflow FMA (R-10017) only, the Controller will proceed to the screen below
after locking; no purge/calibration is necessary.
Steps 8-11 apply to Veriflow FMA set R-10003 only.
8. Hang a saline bag and spike the Saline Bag
9. If only one saline bag is being used, clamp off the flow from the second saline spike to prevent air from entering
the inflow tubing.
10. Connect the waste outflow tube to the waste canister or system.
11. Connect the inflow tube to the sheath (sheath type may vary)
12. Place the sterile Tricera handpiece into the sterile field, pass the handpiece connector plug out of the sterile
field, and plug the Tricera handpiece into the handpiece receptacle (Fig. 1 item 2) on the front panel of the Controller
(Figure 17)
Figure 13
Figure 14
Figure 15
Figure 16
LBL 6028 Rev. K
Page 9 of 306