Table III Guidance and Manufacturer's Declaration – Immunity
ME Equipment and ME Systems that are NOT Life-supporting
The 823190R HAKIM Programmer is intended for use in the electromagnetic environment specified below. The customer or user of the 823190R HAKIM Programmer should
ensure that it is used in such an environment. Deviations from emission and immunity environment for this device may affect its life expectancy.
Immunity Test
IEC 60601 Test Level
3 Vrms
Conducted RF
150 kHz to 80 MHz
IEC 61000-4-6
3 V/m
Radiated RF
80 MHz to 2.5 GHz
IEC 61000-4-3
Table IV Recommended Separation Distances between portable and mobile RF Communications equipment and the 823190R HAKIM Programmer
ME Equipment and ME Systems that are NOT Life-supporting
Recommended Separation Distances for the 823190R HAKIM Programmer
The 823190R HAKIM Programmer is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The customer or user of the 823190R
HAKIM Programmer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the
823190R HAKIM Programmer as recommended below, according to the maximum output power of the communications equipment. Deviations from emission and immunity
environment for this device may affect its life expectancy.
Max Output Power (Watts)
0.01
0.1
1
10
100
Compliance Level
3)=3Vrms
(3)=3V/m
Separation (m)
150 kHz to 80 MHz
D=(3.5/V1)
0.116667
0.368932
1.166667
3.689324
11.66667
Electromagnetic Environment – Guidance
Portable and mobile communications equipment should be separated from
the 823190R HAKIM Programmer by no less than the distances calculated/
listed below:
D=(3,5/V1)
150 kHz to 80 MHz
D=(3.5/E1)
80 to 800 MHz
D=(7/E1)
800 MHz to 2.5 GHz
where P is the max power in watts and D is the recommended separation
distance in meters.
Field strengths from fixed transmitters, as determined by an electromagnetic
site survey, should be less than the compliance levels (V1 and E1).
Interference may occur in the vicinity of equipment containing a transmitter.
Separation (m)
80 to 800 MHz
D=(3.5/E1)
0.116667
0.368932
1.166667
3.689324
11.66667
7
Separation (m)
800 MHz to 2.5 GHz
D=(7/E1)
0.233333
0.737865
2.333333
7.378648
23.33333