2.3. aScope 5 Cysto HD parts
2
1
No.
Part
1
Handle
2
Control lever
Working channel entry
3
Working channel
4
Insertion cord
5
Bending section
6
Distal end
4-5-6
Insertion portion
Connector on the
7
aScope 5 Cysto HD cable
8
aScope 5 Cysto HD cable
9
Protection pipe
3. Use of aScope 5 Cysto HD
Numbers in gray circles below refer to illustrations on page 2.
3.1. Preparation and inspection of aScope 5 Cysto HD
Visual Inspection of the aScope 5 Cysto HD 1
1. Check that the pouch seal is intact before opening. Discard the aScope 5 Cysto HD
if the pouch seal has been damaged, or the expiry date has been exceeded 1 .
2. Make sure to remove the protection pipe from the insertion cord.
3. Check that there are no impurities on or damage to the aScope 5 Cysto HD such as rough
surfaces, sharp edges or protrusions which may harm the patient 2 .
4. Check the deflection of the controllable portion, by moving the control lever on the handle
with your thumb to bend the bending section as much as possible. Then slide the control
lever slowly to its neutral position. Confirm that the bending section functions smoothly
and correctly and returns to a neutral position 5 .
Discard the aScope 5 Cysto HD if it fails any of the check points mentioned above.
Refer to the Ambu displaying units' Instructions for Use for preparing and turning the
Ambu displaying units on 3 .
3
Function
Suitable for left and right hand.
Moves the distal end up or down in a single plane.
Allows for instillation of fluids and insertion
of endoscopic instruments.
Can be used for instillation of fluids and insertion
of endoscopic instruments.
Flexible insertion cord.
Maneuverable part.
Contains the camera, light source (two LEDs), as well as
the working channel exit.
The combination of the insertion cord, bending section,
and distal end.
Connects to the grey socket on Ambu displaying units.
Transmits the image signal to Ambu displaying units.
Protects the insertion cord during transport
and storage. Remove before use.