Table 1. Front Panel Controls - Johnson & Johnson DePuy Synthes FMS VUE II Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 73
No.
Description
Transparent safety doors
1
Doors must be closed for pump operation .
IRRIGATION roller pump
2
Provides fluid inflow .
PRESSURE adjustment pads (mmHg)
3
Controls the pressure level in increments of 5 mmHg .
PRESSURE display (mmHg)
4
Displays both the preset pressure when the base pressure is being set, and the dynamic
pressure following the release of the pressure adjustment pads .
FLOW+ Suction setting
5
Sets the flow rate through the cannula .
SHAVER speed display (RPM)
6
Indicates the preset handpiece speed and direction of rotation .
SHAVER Speed Selection pads (RPM)
7
Increases and decreases handpiece speed .
SUCTION roller pump
8
Provides fluid outflow .
SUCTION Auto-locking slot
9
Locks the tubing in place around the SUCTION roller pump .
RFID antenna
10
Detects correct pump tubing .
PINCH VALVE
11
Moves automatically when the handpiece rotates and selects the appropriate suction
tube .
SHAVER Suction setting
12
Increases and decreases handpiece suction .

Table 1. Front Panel Controls

16

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido