1 . Remove the Inflow
tubing from the package
and close three (3)
white clamps on the
Inflow tubing .
CAUTION: Make sure the Quick Connect port is open before inserting the Quick Connect
tube. To open, press down on the lever at the top of the port.
5 . Connect the pressure sensor Quick Connect tube to the Fill Chamber port on the FMS VUE
II pump . The connector clicks when it is locked .
6 . Position the Fill Chamber in the bracket below the pressure sensor connection (Figure 7) .
Make sure the white clamp on the pressure chamber is closed .
7 . Close the pump door .
4
Figure 7. Position Fill Chamber
MITEK SPORTS MEDICINE
COMPANIES OF
2 .
Lift the door and position
the
green
autolock on
the tube in the
autolock slot on the
pump .
1
4
2
3
26
3 .
Place and center the tube
around the roller .
green
4 . Pull gently on the tube
to engage the tube fitting
opposite the auto-lock while
maintaining tension on the
tube .
1
Quick Connect Tube in port
2
Fill Chamber Bracket
3
Fill Chamber Symbol
4
Quick Connect Port
WARNING! Do not invert Fill
Chamber at any time.
CAUTION! When placing the Fill Chamber
in the bracket, make sure that the Fill
Chamber symbol (3) is facing toward the
user and tubing is not twisted. Improper
positioning could cause the tube to twist
and obstruct the flow of fluid
MITEK SPORTS MEDICINE
COMPANIES OF