Querying The Prosthesis Status; Additional Operating States (Modes); Empty Battery Mode; Overheating Mode - Ottobock 1B1-2 Meridium Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 173
→ The navigation menu opens.
2) In the navigation menu, tap the entry "Functions".
3) Tap the entry "Deactivate Bluetooth".
4) Follow the on-screen instructions.
Switching on Bluetooth
1) Rotate the component or connect/disconnect the battery charger.
→ Bluetooth is switched on for approx. 2 minutes. During this time, the app must be started
in order to establish a connection to the component.
2) Follow the on-screen instructions.
→ If Bluetooth is switched on, the

8.6 Querying the prosthesis status

8.6.1 Query status through cockpit app
1) When the component is connected, tap the  symbol in the main menu.
2) In the navigation menu, tap the entry "Status".
8.6.2 Status display in the cockpit app
Menu option
Trip: 1747
Step: 1747
Batt.: 68

9 Additional operating states (modes)

9.1 Empty battery mode

Beeps and vibration signals are emitted if the available battery charge level drops to 0% (see
Page 79). During this time, damping settings are set to their safety mode values. The prosthesis is
then switched off. You can switch back to basic mode (mode 1) from empty battery mode by
charging the product.
9.2 Mode for charging the prosthesis
The ankle joint of the prosthetic foot is locked during the charging process.
9.3 Safety mode
The product automatically switches to safety mode if a critical system error occurs (e.g. failure of
a sensor signal) or when the battery is empty. Safety mode remains in effect until the error has
been rectified.
Default damping values are activated in safety mode. This makes limited walking possible for the
user even though the system is not active.
The switch to safety mode is indicated by beeps and vibration signals immediately prior to switch­
ing (see Page 79).
Safety mode can be disabled by connecting then disconnecting the battery charger. If the
product switches into safety mode again, this means a permanent error exists. The product must
be inspected by an authorised Ottobock Service Centre.

9.4 Overheating mode

To prevent overheating of the hydraulic unit due to uninterrupted, increased activity (e.g. exten­
ded walking downhill), the range of motion in the ankle joint is restricted as the temperature rises.
Depending on the temperature, this restriction may also result in the ankle joint being locked com­
 symbol appears on the screen.
Description
Daily step counter
Total step counter
Current prosthesis charge
level, as a percentage
Possible actions
Reset the counter by tapping
the "Reset" button.
For informational purposes
only
For informational purposes
only
73

Publicidad

Tabla de contenido
loading