Record Setup - Gima CMS8000 Manual Del Uso

Ocultar thumbs Ver también para CMS8000:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3.4.8 record setup

Select the [RECORD] in the "SYSTEM SETUP" menu to call up the following menu:
Figure 3-19 Record Setup
In this menu, the user can set up to output two waveforms. The waveforms that can be selected include:
ECG1-ECG2
Two ECG waveforms on the screen (If no ECG waveform is currently displayed on the screen,
this item cannot be picked).
SpO
SpO
2
2
cannot be picked).
RESP
RESP waveform (If no RESP waveform is currently displayed on the screen, this item cannot
be picked).
CO2
Displayed waveform either of anesthetic or gas or generated by CO2 module.
IBP1
The first IBP waveform on the screen.(If no IBP waveform is currently displayed on the
screen, this item cannont be picked)
IBP2
The second IBP waveform on the screen(If less than two IBP waveforms)
OFF
No display for this waveform.
• RT REC TIME this item has two options, CONTINUAL and 8s. "CONTINUAL" means once pushing the
"REC/STOP" button on the recorder panel or the monitor panel, the recorder will continuously print out the
waveform or parameter until this button is pushed again.
• TIMING REC TIME OFF used to set up the time interval between two recordings. 10 selections are availa-
ble: "OFF, 10min, 20min, 30min, 40min, 50min, 1hour, 2hours, 3hours and 4hours". The system will start
the recording process according to the selected time interval. The recording time is always 8 seconds.
Note
• RT REC takes priority over TIMING REC.
• REC RATE: this item has two options, 25.0 and 50.0 mm/s.
• REC GRID: used to decide output format: OFF is without grid, and ON is with grid.
• CLEAR REC TASK: used to clear the alarm event that has been generated and is waiting for recording out.
• If two same waveforms are selected, the system will automatically change one of the waveform to a
different one.
3.4.9 event Setup
The monitor has four types of events. You can specify their representations by yourself.
Select the [MARK EVENT] item in the "SYSTEM SETUP" to call up the following menu:
Plethysmogram.(If no SpO
waveform is currently displayed on the screen, this item
2
22