The relation between the partial pressure and the CO2 concentration is given below:
CO2 partial pressure (mmHg) =CO2 concentration (%) × Pamp (ambient pressure, mmHg)
Of CO2 MainStream and CO2 SideStream modules, whichever is selected by the user, autorun measuring
mode is adopted.
16.2.2 operation for Co2 Measurement
• Sidestream
Figure 16-1 Sample Canula
Figure 16-2 SideStream Module
1. The Sidestream module can be used for intubated and non-intubated patient.
2. Insert LEMO interface to socket( CO2) of the monitor
3. Connect the sample cell to the module,the sampling pump automatically turns on.Perform sample cell
zero.
4. Place the sampling line on the patient. Removal the sample cell automatically turns sampling pump off.
5. Capnogram displayed in less than 20 seconds, at an ambient temperature of 25°C, full specifications
within 2 minutes.
• Mainstream
Figure 16-3 MainStream Module
1. The mainstream module can only be used for intubated patient.
2. Set the "WORK MODE" to "MEASUREMENT"in the submenu "OTHER SET" of "CO2 MENU".
3. Connect the module to the monitor.
4. Place the module above on the adapter.
5. Perform airway adapter zero.
6. Connect the airway adapter into the patient's airway.
7. Capnogram, displayed in less than 15 seconds at an ambient temperature of 25ºC, full specifications
within 2 minutes.