79
TEMP alarm limits:
Max. TEMP HI
T1, T2
50
TD
50
• UNIT
To set temperature unit (ºC or ºF).
• DEFAULT
Pick this item to access the TEMP DEFAULT CONFIG dialog box, in which the user may
select whether the FACTORY DEFAULT CONFIG or the USER DEFAULT CONFIG is to be
used. After selecting any of the items and exiting the dialog box, the system will pop up the
dialog box asking for the user's confirmation.

14.3 teMP Alarm message

Among physiological alarms, those belonging to the type that the parameter has exceeded the limits may
activate the recorder to automatically output the parameters and related measured waveforms when the
alarms occur on the condition that the alarm record switch in the related menu is On.
Tables below describe the possible physiological alarms, technical alarms and prompt messages occurring
during TEMP measurement.
Physiological alarms:
Message
T1 TOO HIGH
TI TOO LOW
T2 TOO HIGH
T2 TOO LOW
TD TOO HIGH
Technical alarms:
Alarm Message
T1 SENSOR OFF
T2 SENSOR OFF
T1 ALM LMT ERR
T2 ALM LMT ERR
TD ALM LMT ERR
Prompt message:
Message
T1 EXCEED
T2 EXCEED
Cause
Measuring value of channel 1 is above upper alarm limit.
Measuring value of channel 1 is below lower alarm limit.
Measuring value of channel 2 is above upper alarm limit.
Measuring value of channel 2 is below lower alarm limit.
Difference between two channels is larger than upper limit.
Cause
Temperature cable of channel 1
may be disconnected from the
monitor.
Temperature cable of channel 2
may be disconnected from the
monitor.
Functional safety failure
Functional safety failure
Functional safety failure
Cause
Measuring value of channel 1 is beyond measuring range.
Measuring value of channel 2 is beyond measuring range.
Min. TEMP LO
0
0
Alarm
Remedy
Level
LOW
Make sure that the cable is properly
connected.
LOW
Make sure that the cable is properly
connected.
HIGH
Stop using alarming function of TEMP
module, notify biomedical engineer or
Our service staff.
HIGH
Stop using alarming function of TEMP
module, notify biomedical engineer or
Our service staff.
HIGH
Stop using alarming function of TEMP
module, notify biomedical engineer or
Our service staff.
Step
0.1
0.1
Alarm Level
User-selectable
Alarm Level
HIGH
HIGH