11.6 eCG Alarm Information and Prompt

11.6.1 Alarm Message

Alarms occurring in the process of ECG measurement contain two types: physiological alarm and technical
alarm. Prompt message may also appear in the mean time. For the audio and visual features during the
appearance of these alarms and prompt messages in the process of ECG measurement, please refer to the
related description in Chapter Alarm. In the screen, physiological alarm messages and the prompt messages
able to trigger alarms (general alerts) all displayed in the alarm area of the monitor while technical alarms and
prompt messages unable to trigger alarms are then displayed in the information area of the monitor.
This section does not describe the content about Arr. and ST analysis.
Among physiological alarms, those belonging to the type that the parameter has exceeded the limits may
activate the recorder to automatically output the parameters and related measured waveforms when the
alarms occur on the condition that the alarm record switch in the related menu is On.
Tables below describe respectively the possible various alarms those may occur during the measurement.
Physiological alarms:
Message
Cause
ECG LOST
No ECG signal of the patient is detected.
HR TOO HIGH
HR measuring value is above the upper alarm limit
HR TOO LOW
HR measuring value is below the lower alarm limit
Technical alarms:
Message
ECG LEAD OFF
ECG V LEAD OFF or
ECG C LEAD OFF
ECG LL LEAD OFF or
ECG F LEAD OFF
ECG LA LEAD OFF or
ECG L LEAD OFF
ECG RA LEAD OFF or
ECG R LEAD OFF
ECG INIT ERR
ECG INIT ERR1
ECG INIT ERR2
ECG INIT ERR3
ECG INIT ERR4
ECG INIT ERR5
ECG INIT ERR6
ECG INIT ERR7
ECG INIT ERR8
ECG COMM STOP
ECG COMM ERR
HR ALM LMT ERR
ECG NOISE
Cause
ECG electrodes fall off the
skin or ECG cables fall
off the monitor.
ECG module failure
Occasional communication
failure
Occasional communication
failure
Functional safety failure
ECG measuring signal
is greatly interfered.
Alarm level
Remedy
LOW
Make sure that all electrodes, leads
and patient cables are properly
connected.
Stop using measuring function
HIGH
provided by ECG module, notifies
biomedical engineer or Our service
staff.
HIGH
If failure persists, notify biomedical
engineer or Our service staff.
HIGH
If failure persists, notify biomedical
engineer or Our service staff.
HIGH
Stop using HR alarm function, notify
biomedical engineer or
Our service staff.
LOW
Make sure the patient is quiet,
the electrodes are properly
connected and AC power system
is well grounded.
Alarm level
HIGH
User-selectable
User-selectable
56