• SYS ALM HI, SYS ALM LOW, MEAN ALM HI, MEAN ALM LO, DIA ALM HI, DIA ALM LO are for the user
to set up the alarm limit for each type of pressure. NIBP alarm is activated when the pressure exceeds set
upper alarm limits or falls below lower alarm limits.
NIBP alarm limits
Adult Mode
SYS
40-270 mmHg
DIA
10-215 mmHg
Mean
20-235 mmHg
Pediatric Mode
SYS
40-200 mmHg
DIA
10-150 mmHg
Mean
20-165 mmHg
Neonatal Mode
SYS
40-135 mmHg
DIA
10-100 mmHg
Mean
20-110 mmHg
• UNIT
Pick this item to set measurement unit. (Option: mmHg or kPa)
• INTERVAL
Interval time for automatic measuring. Available selections: 1/2/3/4/5/10/15/30/60/90/120/240/480/960
minutes. Press START/STOP button on the NIBP module to start the first auto measuring.
Pick MANUAL selection in INTERVAL item to set up the measuring mode to MANUAL.
• INFLATION
Can be set only for some customized module.
• RESET
Restore measurement status.
Pick this item to restore initial settings of the pressure pump.
When the pressure does not work properly and the system fails to give message for the problem, pick this
item to activate self-test procedure, thus restore the system from abnormal performance.
• CONTINUAL
Start continuous measuring.
When this item is picked, the menu will disappear automatically.
• CALIBRATE
Calibrate the cuff pressure reading with a calibrated reference manometer. Pick the CALIBRATE item
to start the calibration and the item will change into STOP CAL, which if picked, the system will stop
calibration.
Warning
The calibration of the NIBP measurement is necessary for every two years (or as frequently as dictated by
your Hospital Procedures Policy). The performance should be checked according to the following details.
Procedure of the Pressure Transducer Calibration:
Replace the cuff of the device with a rigid metal vessel with a capacity of 500 ml ± 5%. Connect a calibrated
reference manometer with an error less than 0.8 mmHg and a ball pump by means of a T-piece connector
and hoses to the pneumatic system. Set the monitor in CALIBRATE mode. Inflate the pneumatic system to 0,
50 and 200 mmHg by ball pump separately. The difference between the indicated pressure of the reference
manometer and the indicated pressure of the monitor will not exceed 3 mmHg. Otherwise, please contact
our customer service.
Figure 13-3 Diagram of NIBP Calibration
74