83
The alarm occurs when the value exceeds the set limits.
IBP alarm limits:
Pressure Label
ART
PA
CVP
RAP
LAP
ICP
P1
P2
IBP Transducer Zero
Press the IBP PRESSURE ZERO button on the IBP SELECT menu to call up IBP PRESSURE ZERO menu as
shown below:
Figure 15-5 IBP PRESSURE ZERO
Note
It is the responsibility of the user to ensure that a zero procedure has recently been done on the transducer,
otherwise there will be no recent, valid zero value for the instrument to use, which may result in inaccurate
measurement results.
Zero Calibration of Transducer
Select CH1, the system will zero IBP1. Select CH2, the system will zero IBP2.
Caution
• Turn off patient stopcock before you start the zero procedure.
• The transducer must be vented to atmospheric pressure before the zero procedure.
• The transducer should be placed at the same height level with the heart, approximately mid-axially line.
• Zero procedure should be performed before starting the monitoring and at least once a day after each
disconnect-and-connect of the cable.
The prompt information related to zero, take CH1 for example.
• "SENSOR OFF, FAIL"
Make sure that transducer is not off, then proceed zeroing.
• "IN DEMO FAIL"
Make sure that the monitor is not in DEMO mode. Contact service technician if necessary.
• "PRESSURE OVER RANGE, FAIL"
Make sure that the stopcock is vented to atmosphere. If the problem persists, contact service technician.
• "PULSATILE PRESSURE, FAIL"
Make sure that the transducer is not attached to the patient and that the stopcock is vented to
atmosphere. If the problem persists, contact service technician.
IBP Calibration
Press the IBP PRESSURE CALIBRATION button on the IBP SELECT menu to call up the IBP PRESSURE
CALIBRATE menu as shown below:
Max. Alarm High (mmHg)
300
300
300
300
300
300
300
300
Min. Alarm Low (mmHg)
-10
-10
-10
-10
-10
-10
-10
-10
Step (mmHg)
1
1
1
1
1
1
1
1